Home/Strumenti/Bilance/Bilance accessori/XA100 – Adapter for Pipettes calibration 100 ml for XA.4Y; XA.4Y.A.I Stazioni di

XA100 – Adapter for Pipettes calibration 100 ml for XA.4Y; XA.4Y.A.I Stazioni di

1.780,00 + IVA

SKU: MNT-G6393

Descrizione

XA100 – Adapter for Pipettes calibration 100 ml for XA.4Y; XA.4Y.A.I Stazioni di

XA100 pipette calibration adapter enables calibration and/or verification of piston pipettes by means of a gravimetric method using the XA analytical balence.
Set content:
  • base
  • weighing pan
  • weighing vessel
  • evaporation ring
  • glass chamber with an opening for dosing
Evaporation ring is an integral part of the adapter that increases relative humidity inside the weighing chamber. With this evaporation of the liquid during calibration/verification is stopped.
An integrated weighing pan enables centric loading of the vessel in the weighing chamber. The weighing pan and the vessel are housed in a glass chamber featuring an opening allowing the user to dispense liquid through it, and reducing air drafts.
 
Vessel capacity: 100 ml
 
Adapters for pipettes calibration
 
Pipette calibration adapters are intended for calibration and verification of piston pipettes using gravimetric method. Use of the pipette calibration adapter reduces maximum capacity of the balance by the mass of the weighing vessel. The table presents the minimum reguirements in accordance with ISO 8655 standard. It also enables to select suitable adapter for pipette calibration and respective balance version.
 
Selected volume1 of apparatus under test
V
Resolution
mg
Repeatability and linearity
mg
Recommended balance
1 ยตl = V = 10 ยตl 0.001 0.002 MYA 21.5Y XA 6/21.5Y.M.A
XA 6/21.5Y.M
 
10 ยตl < V = 100 ยตl 0.01 0.02 XA 82/220.5Y.A
XA 82/220.5Y
XA 120/250.5Y.A
XA 120/250.5Y
100 ยตl < V = 1000 ยตl 0.1 0.2
1 ml < V = 10 ml 0.1 0.2
Adapter for pipette calibration MY 11 XA 11 XA 17 XA 100 XA 17 XA 100
Maximum vessel capacity 11 ml 11 ml 17 ml 100 ml 17 ml 100 ml
Weighing vessel mass ~10 g ~10 g ~11 g ~65 g ~11 g ~65 g

1 For practical purposes, the nominal volume may be used to choose the balance. [Reference: ISO8655-6:2002, pg. 2] If the standard uncertainty of measurement of the balance is known (e.g. from the balance calibration certificate), it may be used instead of the repeatability and linearity. The standard uncertainty of measurement shall not be more than two to three times the resolution. [Reference: ISO8655-6:2002, pg. 2]  
 

td {padding:10px; }
tr {text-align:center;}
@media (max-width: 800px) {
table {font-size: 9px}
td {padding: 1px 1px}
}

a{color:#737373;}

Titolo

Torna in cima